Pular para o conteúdo principal
GE Healthcare · CTBrightSpeedLightSpeed244634200501544260112009244540

BrightSpeed Select or Delight Select

Symptom

Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.

Error message

Error: 244634 Subsystem State; DCB:up, CCB:up, -TCB:down -, JEDI:up, TGP/ORP:up TCB:DOWN See Long Description Sun Aug 29 11:13:58 2004 Error: 244634 Host: erct Process: fwmgr File: fwmgr utils.c Line: 660 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:up CCB:up TCB:down JEDI:up TGP/ORP:up.

Error: 200501544 scannerGotResource: Failure had ocurred while performing rxif getHWStatus See Long Description for more details Sat Sep 16 18:16:28 2006 Error: 200501544 Host: ls01 Process: scanRx File: ZeusrpGlobalActions.c Line: 792 Function: Data Acquisition : OC Processing scannerGotResource: Failure had ocurred while performing rxif getHWStatus.

Error: 260112009 Emergency Off button has been pressed See LD for full error Fri May 20 15:06:56 2005 Error: 260112009 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr.c Line: 640 Fri May 20 15:06:56 2005 Host : Table Gantry Processor Ermes # : 260112009 Exception Class : Operator Abort Severity : Secondary File : uif_table_tilt_mgr.c Line# : 640 Function : SYSTEM POWER CONTROL : Emergency Stop Scan Type : None/Unknown Scan: 0/0/0 Emergency Off button has been pressed

Error: 244

Applies to

Models
BrightSpeed, LightSpeed
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.