GE Healthcare · MRSigna2249420
MGD downlaod fails due to defective inifiniband board.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2249420 A Monarch critical error was detected in task defaultServer Error: 2249420 Host: GEMR3T Process: diagCoord File: excLog.c Line: 1714 Task ID: 0x6eca750 Time: 3465 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A A Monarch critical error was detected in task defaultServer (0x3A94E20). Additional information in /usr/g/service/log/mxExLg-d47
Applies to
- Models
- Signa
- Product
- 3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion Infiniband Board 5139563 See comments in Resolution Concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.