Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260125034260134601260125037244633260134602

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Cradle unlocked indicator flashes and wont let the table move See Long Discription for Details Cradle unlocked indicator flashes and wont let the table move

Error message

Error: 260125034 Did not reboot TCB within 20sec. See Long Description for complete error Tue Aug 8 21:50:05 2006 Error: 260125034 Host: tgp Process: tgp_subsys File: ta_func_gt.c Line: 735 Tue Aug 8 21:50:04 2006 Host : Table Gantry Processor Ermes # : 260125034 Exception Class : Secondary Severity : Pri/Soft File : ta_func_gt.c Line# : 735 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Did not reboot TCB within 20sec.

Error: 260134601 Cradle drive error See Long Description for complete error. Fri Oct 5 12:45:30 2007 Error: 260134601 Host: tgp Process: tgp_subsys File: gtcb_cmc_drv_req.cxx Line: 1185 Fri Oct 5 12:45:29 2007 Host : Table Sub-System Ermes # : 260134601 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_drv_req.cxx Line# : 1185 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle drive error

Error: 260125037 Table is in shutdown state, cycling power for the table See the LD for the complete error code Error: 260125037 Table is in shutdown state, cycling power for the table Wed May 25 17:35:00 2005 Error: 260125037 Host: tgp Process: tgp_subsys File: ta_func.c Line: 451 Wed May 25 17:35:00 2005

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues