GE Healthcare · CTLightSpeedVCT26000101218418141674
LightSpeed 1.X - GRE
Symptom
Intermittent rotated images - may have no errors in the log. See Long Description for details Intermittent rotated images - may have no errors in the log.. Rotated images in all of the helical series. Rotated Image two.
Error message
Error: 260001012 The Rotating Controller Interface Bus (RCIB) fault line has been asserted See Long Description for the full error code Sat Jun 9 14:37:06 2001 Error: 260001012 Host: OBC Process: HOP File: hop_fault_state.c Line: 137 Function: No System Function Reported Scan: 9553/3/1 Type: Axial Exception Level: Pri/Most Severe Time: 14:37:06:030 Log Series: 366 The Rotating Controller Interface Bus (RCIB) fault line has been asserted. None of the sub-systems have reported a problem. Check nodes or fault line for failure. Error: 184181 Number of offsets acquired does not match estimate of number of offsets generated See Long Description for complete error Wed Nov 16 10:17:50 2005 Error: 184181 Suite: CT27 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.68 Method: No Method Line: 480 Function: XRAY GENERATION : Scan Data Trigger Acquisition Scan: 41674/201/1 Type: Axial Exception Level: Pri/Most Severe Time: 10:17:49:830 Log Series: 755 Number of offsets acquired does not match estimate of number of offsets generated HW ERROR: Possible hardware problem. Offsets generated (estimate): -4904 Difference Between Generated and Acquired: 9810 Test Margin: 12 Triggers
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296854P2 Encoder - used on HSA/CTi/1.X gantries as the Axial Encoder and on Zeus Tables as the Elevation Encoder. See Long Description for part details Part #: 46-296854P2 Part Type: GP CONSUMABLES Part Weight: 0.88 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION ENCODER FRU: Yes Status: Orderable ICV/Transfer Price: 104.55 USD /Each List Price: 815.40 USD /EACH Comments: FLS - ADDED PER PCN 189960. MS4636 4/22/97 & PCN 184587 MS03330 ** FE ASSIST ** Axial encoder HSA/CTi/1.X mounts on Axial Motor. (212001261 8/5/08) Part #: 46-296854P2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Axial Encoder FRU: Yes Status: Orderable Parts Offered as: Current: 46-296854P2 Used: Not Available Exchange: Not Available Comments: FLS - ADDED PER PCN 189960. MS4636 4/22/97 & PCN 184587 MS03330 5/5/97 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE "U" PART IF AVAILABLE. IF NO "U" USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.