Symptom
Images are showing up on the monitor with autoview but not in Image Works. Archiving images and networking of the images are failing.
Error message
Error: 200000437 Image installation failed for image (for some Exam/Series/Image number) Open Long Description for full error varaiations Sat Jul 16 13:11:02 2005 Error: 200000437 Host: hsa2 Process: dms File: InstallImage.cxx Line: 316 Image installation failed for image 7297/1/1. Reason: Response from SdC timeout. Thu Feb 2 13:32:10 2006 Error: 200000437 Host: ct01 Process: dms File: InstallImage.cxx Line: 713 Image installation failed for image 2336/999/1. Reason: DMS internal error occurred. image path = %s Error code: 200000423 dks1d1s0: Media error: Unrecovered data block read error (asc=0x11, asq=0x0), (data byte 389), Block #7931 (12821) [filter /usr/in/klogpp failed: too slow] in /var/adm/SYSLOG See Long Description
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2169940-78UU or 2169940-78 Octane/Octane II Computer 18 GB Disk Drive. See long description Part #: 2169940-78UU Part Type: GP USED Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: OCT2 18GB 10K RPM DRIVE FRU: Not Available Status: Orderable Stan Pearson, Asset Mgmt move to used material program 6-27-05 Part #: 2169940-78 Part Type: GP CONSUMABLES Part Weight: 1.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: OCT2 18GB 10K RPM DRIVE FRU: Yes Status: Orderable Comments ** FE ASSIST ** 2169940-12 is no longer available can use 2169940-72 9GB or a 2169940-78 18GB Hard drives for a CTi or lightspeed QXI Octane systems as Image or OS disk. Will need to use updated LFC per 2282382-100 rev 2 or 2284291-100 rev 8. The 2169940-78 has 5000 images files where normal is 20000 if used as a CTi Image Disk. Refer to Service Note 0711261b. (212001261 8/20/08)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.