Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT210000445200280418

LightSpeed 4.X - GRE

Symptom

Images fail to display and recon shutdown. Images fail to display and recon shutdown

Error message

Error: 210000445 Error: saveItem() failed with status X. See Long Description for complete error. Fri Sep 30 14:59:58 2005 Error: 210000445 Host: ct99 Process: dms File: DMFInstall.cxx Line: 236 Error: saveItem() failed with status -1 error message TdbExceptionNotImplem ented- for image path /usr/g/sdc_image_pool/reserved/IMSsecBin16/E613U0_S 2U0_I25U0.1128106798_69694 Thu Jul 30 08:55:26 2009 Error: 210000445 Host: H0140CT1 Process: dms File: DMFInstall.cxx Line: 450 Error: saveItem() failed with status 0 error message DMFCreateSession() failed for image path /usr/g/sdc_image_pool/reserved/IMSse7Q2TVf/E1009U0_S1U0_I1U0.1248958526_266351

Error: 200280418 Null pointer detected See long description for full error Mon Mar 26 14:56:32 2007 Error: 200280418 Host: hmct Process: examRxDisplay File: /gspnet/blr/component/commondisplay/view/ Line: 173 Null pointer detected

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2368697-3 HEWLETT-PACKARD HP8000 Computer used in Global Recon Engine (GRE) CONSOLE 2368697-3 GRE CONSOLE HP8000 2368697-3 HP8000 COMPUTER USED IN GRE CONSOLE Part #: 2368697-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKST'N PS2 MOUSE--100 PC FRU: Yes Status: Orderable Comments: 3/4/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THISIS THE GOC2 HP 8000 COMPUTER. ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Please contact the OLC for assistance with new Option Keys
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.