GE Healthcare · Nuclear Medicine13006740

Some Bone Scan images were blank after retrieved from CD in eNTEGRA.

Symptom

Some Bone Scan images were blank after retrieved from CD in eNTEGRA. Also see iTrak 13006740 Problem Description: In version 2.5110 and 2.5202 customer archived a whole body bone scan and 2 spot views. When trying to restore the study, annotation only appears on the display screen. The spots view display correctly. Both anterior and posterior views show the time but the counts in the images are equal to 0. Root Cause: 1. If error occurs during reading Pixel data from Database server, this error is not stored and not processed 2. Error is not generated, Pixel data is empty (no counts is registered)

Applies to

Product
Entegra eNTEGRA Workstation eNTEGRA Workstation
Software
Entegra 2.5202 Software Revision: 2.5202 M4 release Kit Part Number: UGP001440 FMI # / Service Note #: FMI 40770 Mandatory on Request Comments: LSW000419 Software CD PKG Prerequisite: 2.0318 or 2.0319
Component
*Software Software Software
Area
Nuclear Medicine

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.