GE Healthcare · CTLightSpeed260003116

LightSpeed 5.X (80/100kW)

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 260003116 Filter drive tracking detected no motion when motion was requested. See long description for sample error. Fri Sep 14 10:15:52 2007 Error: 260003116 Host: orp Process: orp subsys File: ccb drive timers.c Line: 504 Fri Sep 14 10:15:51 2007 Host : Collimator Ermes # : 260003116 Exception Class : Abort Severity : Primary File : ccb drive timers.c Line# : 504 Function : No System Function Reported Scan Type : Helical Scan: 8516/3/20 Filter drive tracking detected no motion when motion was requested. Verify operation of stepper motor and encoder feedback Start Position = 41999um 39697cnt Desired Position = 0um 0cnt Current Motor Position = 40415um 38200cnt Current Encoder Position= 41998um 39696cnt

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues