GE Healthcare · CTBrightSpeedLightSpeed26000314113155260003139
BrightSpeed HPower or Delight HPower
Symptom
Scanning aborts with collimator filter homing problem. Scanning aborts with collimator filter homing problem
Error message
Error: 260003141 Filter homing problem. See the Long Description for the full error code Fri Mar 4 21:22:12 2005 Host : Collimator Ermes # : 260003141 Exception Class : Abort Severity : Primary File : filter_home.c Line# : 147 Function : No System Function Reported Scan Type : None/Unknown Scan: 13155/1/0 Filter homing problem. Error: 260003139 Filter Switch Error. Both Contacts Are Either Open or Closed. See Long Description for complete error. Error: 260003139 Host : Collimator Ermes # : 260003139 Exception Class : Soft Severity : Pri/Soft Filter Switch Error. Both Contacts Are Either Open or Closed. They Should be the inverse of each other Filter Switch NO Contact: Open Filter Switch NC Contact: Open
Applies to
- Models
- BrightSpeed, LightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required