Pular para o conteúdo principal
GE Healthcare · MRSigna224716122471542249431116808

Cannot scan with many PCI and occasional IRF driver function errors due to IRF2 board.

Symptom

Cannot scan with many PCI and occasional IRF driver function errors. See Long Description See errors, below. Error: 2249431 Host: GEMR_OC4 Process: diagCoord File: mgd820PciErr.c Line: 1378 Task ID: 0x3e2b6e0 Time: 116808 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A A PCI error was detected, but message queue send failed to the exception logging task; errno = 0x3D0002. Harrier Err Enable Reg: 0x00000000; Harrier Err Status Reg: 0x00000000. Harrier Err Address Reg: 0x00000000; Harrier Err Attribute Reg: 0x8000. Mon Oct 5 13:36:00 2009 Error: 2247154 Host: GEMR_OC4 Process: diagCoord File: Receiver.cpp Line: 258 Task ID: 0x7effdf0 Time: 1348 Object: APS Exception Class: Unknown Function: RF: RF Transmit The IRF driver function Receiver::rx_set_r1_atten failed to write a new control value to the RRF. Channel=1, Written Value = 0, Returned Value = 6 If this occurs often you may have RRF chassis communication problem Error: 2247161 Host: GEMR_OC4 Process: diagCoord File: Utns.cpp Line: 221 Task ID: 0x7effdf0 Time: 1350 Object: APS Exception Class: Unknown Function: RF: RF Transmit UTNS WRITE CONFIRM ERROR The IRF driver function Utns::utns_set_notch_period failed to write a new control value to the UTNS in the RRF chassis. Channel=1, Written Value = 8, Returned Value = 24 If this occurs often you may have RRF chassis communication problem.

Error message

2247161 The IRF driver function %s failed to write a new control value to the UTNS in the RRF chassis 2247161 ERROR: EM_IRF_DRVR_UTNS_CONFIRM_ERROR UTNS WRITE CONFIRM ERROR The IRF driver function %s failed to write a new control value to the UTNS in the RRF chassis. Channel=%d, Written Value = %d, Returned Value = %d If this occurs often you may have RRF chassis communication problem. Extended Error: The APS wrote a value to a UTNS board then did not read the same value back. The problem may be on the UTNS board, the DIF board, or the backplane pins of the RRF chassis. Possible cause(s): IRF Board, IRF I/O Board, RRF-DIF Board, UTNS Board, GigaLink Fiber Cable

2247154 The IRF driver function Receiver rx_set_r1_atten failed to write a new control value to the RRF. Task ID: 0x7effdf0 Time: 1352 Object: APS Exception Class: Unknown Function: RF: RF Transmit The IRF driver function Receiver::rx_set_r1_atten failed to write a new control value to the RRF. Channel=4, Written Value = 0, Returned Value = -12 If this occurs often you may have RRF chassis communication problem.-> 2247154 Task ID: 0x7effdf0 Time: 1352 Object: APS Exception Class: Unknown Function: RF: RF Transmit The IRF dr

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2358122-2 IRF2 Board
Area
MR

Full cause & step-by-step resolution