Cannot scan with many PCI and occasional IRF driver function errors due to IRF2 board.
Symptom
Cannot scan with many PCI and occasional IRF driver function errors. See Long Description See errors, below. Error: 2249431 Host: GEMR_OC4 Process: diagCoord File: mgd820PciErr.c Line: 1378 Task ID: 0x3e2b6e0 Time: 116808 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A A PCI error was detected, but message queue send failed to the exception logging task; errno = 0x3D0002. Harrier Err Enable Reg: 0x00000000; Harrier Err Status Reg: 0x00000000. Harrier Err Address Reg: 0x00000000; Harrier Err Attribute Reg: 0x8000. Mon Oct 5 13:36:00 2009 Error: 2247154 Host: GEMR_OC4 Process: diagCoord File: Receiver.cpp Line: 258 Task ID: 0x7effdf0 Time: 1348 Object: APS Exception Class: Unknown Function: RF: RF Transmit The IRF driver function Receiver::rx_set_r1_atten failed to write a new control value to the RRF. Channel=1, Written Value = 0, Returned Value = 6 If this occurs often you may have RRF chassis communication problem Error: 2247161 Host: GEMR_OC4 Process: diagCoord File: Utns.cpp Line: 221 Task ID: 0x7effdf0 Time: 1350 Object: APS Exception Class: Unknown Function: RF: RF Transmit UTNS WRITE CONFIRM ERROR The IRF driver function Utns::utns_set_notch_period failed to write a new control value to the UTNS in the RRF chassis. Channel=1, Written Value = 8, Returned Value = 24 If this occurs often you may have RRF chassis communication problem.
Error message
2247161 The IRF driver function %s failed to write a new control value to the UTNS in the RRF chassis 2247161 ERROR: EM_IRF_DRVR_UTNS_CONFIRM_ERROR UTNS WRITE CONFIRM ERROR The IRF driver function %s failed to write a new control value to the UTNS in the RRF chassis. Channel=%d, Written Value = %d, Returned Value = %d If this occurs often you may have RRF chassis communication problem. Extended Error: The APS wrote a value to a UTNS board then did not read the same value back. The problem may be on the UTNS board, the DIF board, or the backplane pins of the RRF chassis. Possible cause(s): IRF Board, IRF I/O Board, RRF-DIF Board, UTNS Board, GigaLink Fiber Cable 2247154 The IRF driver function Receiver rx_set_r1_atten failed to write a new control value to the RRF. Task ID: 0x7effdf0 Time: 1352 Object: APS Exception Class: Unknown Function: RF: RF Transmit The IRF driver function Receiver::rx_set_r1_atten failed to write a new control value to the RRF. Channel=4, Written Value = 0, Returned Value = -12 If this occurs often you may have RRF chassis communication problem.-> 2247154 Task ID: 0x7effdf0 Time: 1352 Object: APS Exception Class: Unknown Function: RF: RF Transmit The IRF dr
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2358122-2 IRF2 Board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.