AMX Tube and CPU need to be proper version due to tube anode cooling
Symptom
AMX Tube and CPU need to be proper version due to tube anode cooling See Long Description 46-264974G6 IS NOT BACKWARDS COMPATIBLE TO 46-264974G5 PER GARY MICHALEK, ENGINEERING THE ORION TUBE COOLS MUCH FASTER THAN THE HRT. IF THE BOARDS WERE EVER SWAPPED, THEY WOULD WORK WITH NO PROBLEM RIGHT UP TO THE POINT WHERE THE HRT TUBE ANODE MELTED OFF THE STATOR, CAUSING AN IMPLSION TO THE INSERT WHICH CAN CATCH THE OIL ON FIRE
Visual identification of HRT-75 verses Orion MX75TH11 tube See Long Description There are two visual differences in determining the type of tube you have on your portable, 1, the Orion high voltage horns are attached radially to the casing verses on the HRT-75 they are top tangential orientation (appear to be laying on top of tube) when viewed from park position, and 2, the end caps are narrower and higher on the Orian
Applies to
- Product
- AMX4+ AMX4
- Component
- CPU Board. Units with Orion X-Ray tubes use CPU p/n 46-264974G6, units with MX75 X-Ray tubes use 46-264974G5 See Long Description CPU Board. Units with Orion X-Ray tubes use CPU p/n 46-264974G6, units with MX75 X-Ray tubes use 46-264974G5 HRT-75 part number 46-155750G207 or D2642P and the Orion MX75TH11 tube is D2652P
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
- AEC Backup Time message
- New style ion chamber does not have a preamp
- Precision 500D Table Fluoro Carriage Front Bumper is not properly adjusted, allows excessive travel toward the front of the table.
- Precision 500D Aurora, dose readout does not update when an overhead Rad Tube Exposure is made, at any technic due to abnormally low mR/mAS from the Rad Overhead XRT.
- P500D Aurora with M4 Software (May 2006 Release), Dose Readout Correct on Console, Dose Printer Prints 0.0 due to a M4 Software Bug