GE Healthcare · X-ray R&F

AMX Tube and CPU need to be proper version due to tube anode cooling

Symptom

AMX Tube and CPU need to be proper version due to tube anode cooling See Long Description 46-264974G6 IS NOT BACKWARDS COMPATIBLE TO 46-264974G5 PER GARY MICHALEK, ENGINEERING THE ORION TUBE COOLS MUCH FASTER THAN THE HRT. IF THE BOARDS WERE EVER SWAPPED, THEY WOULD WORK WITH NO PROBLEM RIGHT UP TO THE POINT WHERE THE HRT TUBE ANODE MELTED OFF THE STATOR, CAUSING AN IMPLSION TO THE INSERT WHICH CAN CATCH THE OIL ON FIRE

Visual identification of HRT-75 verses Orion MX75TH11 tube See Long Description There are two visual differences in determining the type of tube you have on your portable, 1, the Orion high voltage horns are attached radially to the casing verses on the HRT-75 they are top tangential orientation (appear to be laying on top of tube) when viewed from park position, and 2, the end caps are narrower and higher on the Orian

Applies to

Product
AMX4+ AMX4
Component
CPU Board. Units with Orion X-Ray tubes use CPU p/n 46-264974G6, units with MX75 X-Ray tubes use 46-264974G5 See Long Description CPU Board. Units with Orion X-Ray tubes use CPU p/n 46-264974G6, units with MX75 X-Ray tubes use 46-264974G5 HRT-75 part number 46-155750G207 or D2642P and the Orion MX75TH11 tube is D2652P
Area
X-ray R&F

Full cause & step-by-step resolution