GE Healthcare · CTBrightSpeedHiSpeed26011821150018
BrightSpeed HPower, Delight, Velocity or MERI HPower
Symptom
Scanner Hardware Stopped Scan with High Voltage errors
Error message
Error: 260118211 ( 60-0313 )Inverter max. ILR current detected. This is a hardware detection of maximum current in serial resonant circuit. Z6,Z7= 0h,0h See long description for complete error Mon Feb 13 17:22:21 2006 Error: 260118211 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Mon Feb 13 17:22:20 2006 Host : Jedi Ermes # : 260118211 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50018/1/4 (60-0313)Inverter max. ILR current detected. This is a hardware detection of maximum current in serial resonant circuit.
Applies to
- Models
- BrightSpeed, HiSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2214527 Jedi standard HV tank for HiSpeed Xi, HP 60 and BrightSpeed See Long Description for details Part #: 2214527 Part Type: GP CONSUMABLES Part Weight: -3059969.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI STD HV TANK (SPARE) FRU: Yes Status: Orderable Comments: FLS - JEDI STD HV TANK U/O YMS PROSPEED LXI,FXI,DXI REPLACES 2186730 DIFFERENCE IS PACKAGING FOR SERVICE. HQT: QUALITY ISSUE 10/98 ALL TANKS WITH SERIAL NUMBER PRIOR TO 22124BU4 ARE BAD. FE SHOULD CHECK HIS SERIAL # EUROPE PURGING GLOBE WILL SEND NEW MATERIAL AS QUICKLY AS POSSIBLE. AMERICA HAS NO STOCK ONHAND AS OF 10/22/98MS08992 RETURN DEFECTIVES TO: SWAP (MICHELL CHRISTOFFERSON), 2200 E. COLLEGE, BUILDING 11, CUDAHY, WI 53110. ADDED BY MS20544 ON 3/25/02. RETURN DEFECTIVES TO RECYCLING7/23/01 MS5277 TAKEN OFF RECYCLING PROGRAM 8/30/01 MS5277 TAKEN OFF RECYCLING PROGRAM PER L.BENIWAL (MSA8429 09/05/02) GELS - 2214527 = 2186730 + WOODEN BOX 10DISPUTES DEFECTIVE PARTS TO BE PUT ON SHELVES FOR ANALYSIS PDB00/07/27 KANBAN CSS F00,JEDI STD HV TANK (SPARE) 2214527 Jedi standard HV tank for HiSpeed and HP 60 Systems HV Tank for HP60 systems - must be disassembled from the Inverter Power assembly, and the new tank installed. Make certain you follow the disassembly/reassembly instructions in the Service Manuals exactly, as the process is quite difficult, and you are prone to scraped knuckles and possible broken connectors or scraped cables if you are not very careful. Picture of removed HV tank below:
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.