Symptom
While raising table, table suddenly drops a short distance with a loud bang.
Error message
Error: 260115196 Elevation lower limit switch is activated. see long description for details Sat Mar 15 06:13:29 2008 Error: 260115196 Host: tgp Process: tgp subsys File: mtcb emc isrfunc.c Line: 79 Sat Mar 15 06:13:29 2008 Host : Table Sub-System Ermes # : 260115196 Exception Class : Secondary Severity : Pri/Soft File : mtcb emc isrfunc.c Line# : 79 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation lower limit switch is activated. Position is 516886 um from the floor.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2342733-2 H-POWER, AC ELEVATION MOTOR, EMC2 - used on HP table. See Long Description for details. Part #: 2342733-2 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER, AC ELEVATION MOTOR, EMC2 FRU: Yes Status: Orderable Comments: Part added per DST-HP60 16 slice RSPL. 040906 MS13452. 2100671 CT TABLE GAS SPRING - Used on ALL HSA, CTi, LightSpeed 1.X through 5.X tables Part #: 2100671 Part Type: GP CONSUMABLES Part Weight: 1.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: CT Table gas spring FRU: Yes Status: Orderable Comments : FLS - REPLACES 46-327064P1 & 46-296891P1 PER PCN 161843 7-20-93 MS4398. NOT ON SSS: SUSPA IS NOT SUPPLIER. TOOL FOR REPLACING GAS SPRING IS 2144261 ************************************************************PART HAS BEEN DE-LISTED AS A HAZARDOUS MATERIAL ITEM. PART IS NOT HAZARDOUS FOR ANY MODE OF TRANSPORTATION. HAZARDOUS PAPERWORK IS NOT NEEDED. 4/99 ************************************************************ ALSO USED ON QXI LIGHTSPEED ALSO USED ON LIGHTSPEED GELS - "NON FLAMMABLE GAS - HAZ IATA DESC. ""ARTICLES, PRESSURIZED, PNEUMATIC "" UN 3164 - CLASS 2.2",GAS SPRING HAZ #8 is the shock 2343479-2 Table Elevation Linear Actuator , H-PWR 5.x Table See Long Description for details Part #: 2343479-2 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBRRepairable: NoReturnable: NoHazardous Material: NDescription: Actuator, Linear, H-Power based Tables, PinnedFRU: Not AvailableStatus: Orderable Comments: Added per ECR 2007660 as replacement for 2342479. ms13452 041217
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.