GE Healthcare · MRSigna224151722415462241537

Will Not Scan with Gradient Errors Due to Z Output Shelf.

Symptom

Will Not Scan with Gradient Errors.

Error message

2241517 The GIP reported a Transistor Fault on the X Axis of the GPM. This Fault Could be Related to Noise. Tue Apr 20 20:24:32 2004 Error: 2241517 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2218 The GIP reported a Transistor Fault on the X Axis of the GPM. This fault could be related to noise. Possible causes: Cable between the GIP and GPM, X Axis GPM Main Board or Output Shelf.->

2241546 The GIP reported a second Transistor Fault on the % Axis of the GPM. .2241546Host: lx_mr Process: NSPFile: HPC_PORT_ID:GAP/GIP/GP Line: 661The GIP reported a second Transistor Fault on the Z Axis of the GPM.Power cycle the main breaker on GPM Power Supply to recover from this condition. If this fault occurs again, replace the GPM Z Axis Output Shelf.->

2241537 The X Axis of the GPM failed to go to Ready when Commanded. Tue Apr 20 20:24:32 2004 Error: 2241537 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 628 The X Axis of the GPM failed to go to Ready when commanded. Check the cable between the GIP & GPM, and the LED status on the X Axis of the GPM. Guarantee that the switch on the GPM

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2175599-4 TECHRON #101577-2, OUTPUT SHELF Z Part #: 2175599-4Part Type: GP REPAIRABLES Part Weight: 29.54 LBRRepairable: YesReturnable: YesHazardous Material: NDescription: TECHRON #101577-2, OUTPUT SHELF ZFRU: YesStatus: OrderableICV/Transfer Price: 445.02 USD /Each ICV/Transfer Price: 445.02 USD /Each List Price: 26536.30 USD /EACH ======================Parts Offered as:Current: 2175599-4 Harvested: Not AvailableExchange: Not Available Comments: GELS - ****WARNING: THIS PART IS SUBMITTED TO THE 2ND OPINION PROCESS*** AN APPROVER'S NAME SHOULD BE MENTIONED IN C915. PASTOURET/OLCE,Z OUTPUT SHELF (GPM)
Area
MR

Full cause & step-by-step resolution