GE Healthcare · MRSigna224151722415462241537
Will Not Scan with Gradient Errors Due to Z Output Shelf.
Symptom
Will Not Scan with Gradient Errors.
Error message
2241517 The GIP reported a Transistor Fault on the X Axis of the GPM. This Fault Could be Related to Noise. Tue Apr 20 20:24:32 2004 Error: 2241517 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2218 The GIP reported a Transistor Fault on the X Axis of the GPM. This fault could be related to noise. Possible causes: Cable between the GIP and GPM, X Axis GPM Main Board or Output Shelf.-> 2241546 The GIP reported a second Transistor Fault on the % Axis of the GPM. .2241546Host: lx_mr Process: NSPFile: HPC_PORT_ID:GAP/GIP/GP Line: 661The GIP reported a second Transistor Fault on the Z Axis of the GPM.Power cycle the main breaker on GPM Power Supply to recover from this condition. If this fault occurs again, replace the GPM Z Axis Output Shelf.-> 2241537 The X Axis of the GPM failed to go to Ready when Commanded. Tue Apr 20 20:24:32 2004 Error: 2241537 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 628 The X Axis of the GPM failed to go to Ready when commanded. Check the cable between the GIP & GPM, and the LED status on the X Axis of the GPM. Guarantee that the switch on the GPM
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2175599-4 TECHRON #101577-2, OUTPUT SHELF Z Part #: 2175599-4Part Type: GP REPAIRABLES Part Weight: 29.54 LBRRepairable: YesReturnable: YesHazardous Material: NDescription: TECHRON #101577-2, OUTPUT SHELF ZFRU: YesStatus: OrderableICV/Transfer Price: 445.02 USD /Each ICV/Transfer Price: 445.02 USD /Each List Price: 26536.30 USD /EACH ======================Parts Offered as:Current: 2175599-4 Harvested: Not AvailableExchange: Not Available Comments: GELS - ****WARNING: THIS PART IS SUBMITTED TO THE 2ND OPINION PROCESS*** AN APPROVER'S NAME SHOULD BE MENTIONED IN C915. PASTOURET/OLCE,Z OUTPUT SHELF (GPM)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.