GE Healthcare · CTLightSpeedVCT260132918
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Can't perform fastcal. Detector Rail Temperature Out Of Range
Error message
Error: 260132918 Rail Heater Has Opened. Heater: Rail Low or High Channel See long description for details Sun May 13 18:41:14 2007 Error: 260132918 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 1664 Sun May 13 18:40:32 2007 Host : Das Ermes # : 260132918 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 1664 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Rail Heater Has Opened Heater: Rail Low or High Channel
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5150070 or 5212666 DHC Detector Heater Controller - used on VCT and HD750 See LD for details 5150070 is non-RHOS complaint 5212666 is RHOS Complaint Part #: 5212666 RHOS compliant Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: DHC Assembly FRU: Yes Status: Orderable Part #: 5150070 non-RHOS compliant Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: DHC ASM FRU: Yes Status: Non Orderable <as of July 4, 2006 Comments: Added per ECR 2030869 6/26/2006, MS11892. DHC Detector Heater Controller
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.