Pular para o conteúdo principal
GE Healthcare · MRSigna2223854

TPS is Resetting Automatically When they Arrive on Site Due to Overnight Diagnostics.

Symptom

Customer Reporting That the TPS is Resetting Automatically When They Arrive on Site. Tuesday Mornings Only. This does not occur on other days of the week.

Error message

2223854 Extended TPS Reset in progress (up to 8 minutes). Please be patient.-> See Long Description 2223854 ERROR: EM_mrNSP_VRE_RESET_PROLONGED Extended TPS Reset in progress (up to 8 minutes). Please be patient. Extended Error: One or more ICNs of the VRE might have been non-responsive during a TPS Reset. We therefore reboot all ICNs in an attempt to recover. The operator will experience longer TPS Reset times because of this.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues