Pular para o conteúdo principal
GE Healthcare · MRSigna22238492252109

Application Software does not boot due to defective AGP Board.

Symptom

Application Software does not complete the boot process with TPS not responding message

Error message

2223849 AGP EPROM power-up boot failed See Long Description Error: 2223849 Host: rmns Process: NSP File: TardisReset.cc Line: 1781 AGP EPROM power-up boot failed. Communication Timed-out from AGP board. Verify that MCP820 boards are installed and seated properly for APS and AGP. Extended Error : MCP820 boards can be identified by the word MCP820 printed on the board extraction lever.

2252109 Power up test found the following errors: AGP power-up diagnostics failed PHB/PPCW test See long description. Error 2252109 Power up test found the following errors: AGP power-up diagnostics failed PHB/PPCW test

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2294300-16 HDMR2 AGP-APS motorola CPCI6020 circuit board 2294300-16 HDMR2 AGP-APS motorola CPCI6020 circuit board
Area
MR