GE Healthcare · CTLightSpeedVCT200002368230020203
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner shuts down on its own during scanning. Scanner shuts down on its own during scanning
Error message
Error: 200002368 A set of processes are deadlocked waiting for each other to start. This deadlock has caused the system to shutdown! See Long Description for full error. Mon Oct 9 07:55:18 2006 Error: 200002368 Host: ct01 Process: cupMonitor File: ssProcStart.c Line: 997 A set of processes are deadlocked waiting for each other to start. "afQueueMgr, examRxDisplayMgr, sfm_archive, scanmgr, scanfilemgr, headermgr, coordinator, rtsmgr, reconSelftest, queuemgr, procICE, accs, vavQueueMgr, stcIf, sdm, scanRx, retroRecon, reconMgmt, readcfgServer, ppsproxy, piRxMgmt, mtran, examRxEnv, dms, visScript, ces, awStart, WLServer, ObjMgr" are the likely culprits. This deadlock has caused the system to shutdown! Error:230020203 DARC applications were not shut down from a previous startup. This is most likely due to an abnormal shutdown and/or reboot on the OC computer See Long description for error Tue Jul 21 22:50:49 2009 Error:230020203 Host: darc Process: darc init File: darc init.cxx Line: 148 DARC applications were not shut down from a previous startup. This is most likely due to an abnormal shutdown and/or reboot on the OC computer.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5183547-38 HP8400 with 4GB Memory See Long Discription for Details Part #: 5183547-38 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: GEHC MI and CT Integrated Buy Level PC Workstation. HP8400 with 4GB Memory FRU: Yes Status: Orderable Comments: Per ECR 2050682 & GOC6 RSPL, 212008898, 7/14/08. ** FE ASSIST ** This part is not qualified for PET systems, CT only This part requires a Software Load from Cold LFC during replacement. For all CT and RT Products, verify you have the latest software and Service Pack for your system before trying to install this part. The software versions listed below are a reference, refer to the latest revision of the software Load Cold Document for your system for more information. For 4,8,16 slice CT HPower and RT4 slice systems, order 07MW11.10, For 4,8,16 slice CT Brightspeed systems, order 07BW08.4 For 16 slice RT systems, 08BW20.4 For 32, 64 slice GOC5 based systems, order 07MW18.4, For 32, 64 slice GOC6 based systems, order 08MW33.2 Future versions of the software releases noted above will are also compatible with this part. (212001261 10/21/08)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.