Pular para o conteúdo principal
GE Healthcare · MRSigna222466622245802224725

EPT Fails to Run the B-zero Dither Test in the Body Mode Only Due to Plastic Alignment Tabs on the Grafidy Base plate.

Symptom

EPT Fails to Run the B-zero Dither Test in the Body Mode Only. A full pass of EPT selected and the White Pixel tests complete fine for both head and body. Calibration halts when the Body B-zero Dither test begins.

Error message

2224666 Auto Prescan failed, signal too small. Please try Manual Prescan. Fri Jan 23 14:39:55 2004 2224666 lx-mr psc PscMainControl.cc 1271 Auto Prescan failed, signal too small. Please try Manual Prescan.

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

2224725 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan) See Long Description Error: 2224725 Host: hopgemr1a Process: NSP File: APS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan).

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution