GE Healthcare · MRSigna2218725
ROTMI Errors Due to HDxt MR Software release 15.0_0910a.
Symptom
Error Occurred During Geometry Calculations Error occurred during Geometry Calculations. Re-prescribe slices/slab in Graphic Rx Screen.
Error message
2218725 Error occurred during Geometry Calculations. Re-prescribe slices/slab in Graphic Rx Screen. 2218725 scn RxManagerIF.m 918 Error occurred during Geometry Calculations. Re-prescribe slices/slab in Graphic Rx Screen.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.