Upgraded System Does Not Recognize Legacy Head Coil due to LPCA Configuration Type Setting Incorrect in Guided Install.
Symptom
Upgraded System Does Not Recognize Legacy Head Coil See Long Description This system was just upgraded. When the FE plugs the Quad Head Coil into the legacy port on the LPCA, a message pops up as shown below, and he cannot scan with this coil. This message is not logged. Use Auto coil config to install HNS coil, do not use CFM to config this coil
Error message
Use Auto coil config to install HNS coil, do not use CFM to config this coil See Long Description This message appears on the screen when plugging in a coil. It is not logged.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.