VIBRANT Cal Scan Fails with DP Server Error due to Software problem.

Symptom

VIBRANT Cal Scan Fails with DP Server Error See Long Description and Attachment iTrak # 13273925 and # 13259484. SPR # MRIhc43267 Software on site was upgraded from 14.0 M5A to 15.0 M4 on evening of June 17, 2009. Next patient scans were following morning and on the first Breast exam, the system stopped scanning at the calibration scan for the Vibrant series. TPS reset was performed to recover. Later in the day, a second Breast exam began and encountered the same issue - scanner stopped with failed prescan & communication errors when attempting cal scan for the Vibrant series. A TPS reset again was the recovery method. See attachment for a full accounting of the logged errors.

Error message

2223621 Internal Software Error. Can't callback on undefined NMRID:CM_ECHO_TPS. 2223621 Exception Class: Unknown Function: Software/Firmware: Host Communications Internal Software Error. Can't callback on undefined NMRID:CM_ECHO_TPS.

2223625 Send packet opcode:8200 seqnum:2663 rev:2 type:1 length:4 req:NSP resp:CM_ECHO_TPS status:0 failed. 2223625 Exception Class: Unknown Function: Software/Firmware: Host Communications Send packet opcode:8200 seqnum:2663 rev:2 type:1 length:4 req:NSP resp:CM_ECHO_TPS status:0 failed.

2248321 Communication Error. The callback for SCSM state 0x4 was not received in time. The DP server is not responding, the server might be hung. Please perform a TPS reset to recover.-> 2248321 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: VIBRANT Coil: 8Brst BL SMI Scan: 15642/6 Communication Error. The callback for SCSM state 0x4 was not received in time. The DP server is not responding, the server might be hung. Please perform a TPS reset to recover.->

2218592 The MrMail send of Ready to TPS timed out

2224666 Auto Prescan failed, Pulse generation failed. Probable PSD software error. Pulse generation timed out. It may b

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.