GE Healthcare · CTLightSpeed18210518212415472511547058
LightSpeed 4.X - GOC2
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 182105 LAN Watchdog TIMEOUT for uuid 93 (193=OBC; 93=ETC). See LD for error example. Tue Oct 25 22:34:14 2005 182105 3 CT01 STC Master Exc Handler exm_lan_wdog.c No Method 793 Function: No System Function Reported Scan: 161/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 22:33:56:310 Log Series: 223 LAN Watchdog TIMEOUT for uuid 93 (193=OBC; 93=ETC). HW/SW Error: Did not receive LAN Watchdog message. Will need a Hardware Reset. Error: 182124 LAN Watchdog TIMEOUT for uuid 93 (193=OBC; 93=ETC). See LD for error example. Tue Oct 25 22:34:14 2005 182124 3 CT01 STC Master Exc Handler exm_lan_wdog.c No Method 799 Function: No System Function Reported Scan: 161/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 22:33:55:255 Log Series: 223 LAN Watchdog TIMEOUT for uuid 93 (193=OBC; 93=ETC). HW/SW Error: Did not receive LAN Watchdog message. (Last Wdog Probe Sent:1547251 Last Response Rcvd:1547058 (in Ticks)).
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2377227 UTP Transceiver or Media Convertor used on OC with 1X - 4X Gantry - replaces 2148559 10BASE-2 TO UTP CONVERTER See Long Description for Details. 2377227 UTP Transceiver or Media Covnerter for GRE GOC3 Consoles with HP-8000 CPU and older style Gantries LS1X - 4X Part #: 2377227 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: UTP TRANSCEIVER - BNC TO RJ45 With I.T.E PWR SPPLY 24000027 FRU: Yes Status: Orderable Part #: 2148559 Part Type: GP CONSUMABLES Part Weight: 1.36 LBR Repairable: No Returnable: No Hazardous Material: N Description: 10BASE-2 TO UTP CONVERTER FRU: Yes Status: Orderable Comments: ***OBSOLETE PART Case # 11345706*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-49 ** FE ASSIST ** For P/N 2148559 - Please order the newer media converter 2377227 and external terminator 2154019. The 2154019 50 ohm external terminator is required.(212001261 10/2/07) I re addded my note 7/28/08 as someone removed it 2154019 50 OHM BNC GROUNDED TERMINATOR See Long Description for Detail Part #: 2154019 Part Type: GP CONSUMABLES Part Weight: 0.02 LBR Repairable: No Returnable: No Hazardous Material: N Description: 50 OHM BNC GROUNDED TERMINATOR FRU: Yes Status: Orderable Comments: FLS - 50 OHM BNC GROUNDED TERMINATOR U/O CT/I AND HELIOS 2197229-5 ETHERNET SWITCH, 8 PORT used in 1X/2X/3X/4X GOC1/GOC2 See long description for more details Part #: 2197229-5 Part Type: GP CONSUMABLES Part Weight: 1.94 LBR Repairable: No Returnable: No Hazardous Material: N Description: ETHERNET SWITCH, 8 PORT FRU: Yes Status: Orderable Comments: FLS - ETHERNET SWITCH, 8 PORT. ADDEDPER PCN 220607. REPLACES P/N 2197229-3 AND -4. USE EXISTINGSTOCK FIRST. MS20544, 8/15/02.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.