GE Healthcare · CTLightSpeed89600643527743
Images are not displayed in the autoview port during patient scanning.
Symptom
Images are not displayed in the autoview port during patient scanning. See Long Description 5 years of Images are not displayed in the autoview port during patient scanning
Error message
From the var/adm more messages -> SCSI Error: (1:2:0) Status=02h (CHECK CONDITION) - Key=3h (MEDIUM ERROR); FRU=D2h - See LD for more information SCSI errors point to scsi id 2 in /var/adm/messages Oct 17 07:10:23 bhmgect4 kernel: SCSI Error: (1:2:0) Status=02h (CHECK CONDITION) Oct 17 07:10:23 bhmgect4 kernel: Key=3h (MEDIUM ERROR) ; FRU=D2h Oct 17 07:10:23 bhmgect4 kernel: ASC/ASCQ=16h/00h "" Oct 17 07:10:23 bhmgect4 kernel: CDB: 28 00 00 88 B8 3F 00 00 08 00 Oct 17 07:10:23 bhmgect4 kernel: Oct 17 07:10:29 bhmgect4 kernel: SCSI Error: (1:2:0) Status=02h (CHECK CONDITION) Oct 17 07:10:29 bhmgect4 kernel: Key=3h (MEDIUM ERROR); FRU=D2h Oct 17 07:10:29 bhmgect4 kernel: ASC/ASCQ=16h/00h "" Oct 17 07:10:29 bhmgect4 kernel: CDB: 28 00 00 88 B8 3F 00 00 08 00 Oct 17 07:10:30 bhmgect4 kernel: scsi3: ERROR on channel 0, id 2, lun 0, CDB: Read (10) 00 00 88 b8 3f 00 Oct 17 07:10:30 bhmgect4 kernel: Info fld=0x88b840, Current sdd: sense key Medium Error Oct 17 07:10:30 bhmgect4 kernel: Additional sense: Data synchronization mark error Oct 17 07:10:30 bhmgect4 kernel: end_request: I/O error, dev sdd, sector 8960064 Oct 17 07:10:30 bhmgect4 kernel: I/O error in filesystem ("md0") meta-data dApplies to
- Models
- LightSpeed
- Product
- LightSpeed RT4 WB O G4 G HPT WB O M4 G HPT LightSpeed RT4 with GRE and MDAS4 or GDAS4 gantry At GE Medical Systems, we’re pioneering exciting CT technology for Radiation Therapy planning. Introducing the newest addition to our LightSpeed® series – LightSpeed RT, the first multi-slice, wide-bore CT scanner with the largest field of view in the industry. With a generous 80 cm gantry opening, the LightSpeed RT bore is 10 cm larger than conventional scanners. This provides complete flexibility for patient positioning freedom – even for the most demanding oncology needs. And the 65 cm variable field of view delivers 15 additional centimeters of visual coverage, allowing clinicians to see a patient’s entire anatomy for precise radiation therapy planning. Position and precision Thin, multi-slice imaging is the extra bonus – enabling respiratory gating, providing precise target delineation and delivering higher resolution digitally reconstructed radiographs (DRRs) than ever before. All adding up to the ability for you to deliver what really matters most – better cancer care for your patients. Plus, the LightSpeed RT scanner is available with the Exact™ couch insert, enabling you to consistently replicate patient immobilization for treatment. With LightSpeed RT, the future of your Radiation Oncology department is wide open. Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5117866-5 73 GB HDD used in HPxw8200 GRE. See Long Discription for Details. Part #: 5117866-5 Part Type: GP CONSUMABLES Part Weight: -139998.60 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73 GB HDD for Lighting System Controller 5117866-2 FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.