GE Healthcare · MRSigna22448512244577224457822500832218527
Current Distortion Fault Due to SGA Control Board.
Symptom
Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.
Error message
2244851 The GP is reporting a Gradient Amplifier %c Axis Current Distortion Fault. the same error as 2244577 on 11X, G3, openMR4 and 2244578 on 12X, E2, 14M4 (HDx, HDe), and openMR5. See Long Description 2244851 error code on Post 14M5 and 12.0 M5, is the same error as 2244577 on 11X, G3, openMR4and 2244578 on 12X, E2, 14M4 (HDx, HDe), and openMR5.2244851 The GP is reporting a Gradient Amplifier %c Axis Current Distortion Fault Example error message (could be any gradient axis, Y-axis is shown in this example): Error: 2244851 Host: lx-mr2 Process: NSP File: SCP:Grad Line: 2851 The GP is reporting a Gradient Amplifier Y Axis Current Distortion Fault.-> Extended error message This is a generic fault message indicating that the Gradient Amplifiers are not accurately following the current command from the IPG/SRF. If other faults occurred concurrently, they will provide useful information to assist with troubleshooting. Else, swap the control cable between the GP and the failing Gradient Amplifier with the Gradient Amplifier on another axis (swap one end of the cable only). If the fault moves, replace the Gradient Amplifier or Gradient Amplifier Control Board (board can be swapped with
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2250375-4 CONTROL_BOARD ASSEMBLY_SGA 2250375-4 CONTROL_BOARD ASSEMBLY_SGA 2250375-4 Part Desc: CONTROL_BOARD ASSEMBLY_SGA
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.