Application Software does not complete the boot process with TPS Download failure due to Dual PPC Module.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2227556 *** Reset/Download TPS was not successful. *** 2227556 sdm sdm_main.c 320 *** Reset/Download TPS was not successful. *** 2219628 The BAM Check Packet failed to reach the TPS Tue Jun 15 08:58:55 2004 Error: 2219628 Host: IRIS Process: scn File: SrxIPScreen.m Line: 1956 The BAM Check Packet failed to reach the TPS.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- SecondOpinion Dual PPC Module 2222532-3 See comments in resolution concept for second opinion requirements before ordering this part See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.