Pular para o conteúdo principal
GE Healthcare · MRSigna22415522241522

CWA. Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2241552 The GIP Reported Several PPBM Faults on the Z Axis of the GPM. Mon Apr 26 19:56:01 2004 Error: 2241552 Host: rmnf Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2189 The GIP reported several PPBM Faults on the Z Axis of the GPM. Power cycle the main breaker on the GPM Power Supply to recover from this condition. If this fault occurs again, check the GPM & the GPM Power Supply.

2241522 The GIP reported a PPBM Fault on the Y Axis of the GPM. The GPM will recover when the Supply voltage changes to a safe range. Possible causes: Y Axis GPM PPBM, Main Board, Output Shelf, or the GPM Power Supply. Extended Error: This error is generated by the PPBM and would occur under the following conditions a) PPBM was Overtemp b) Input voltage from GPM Power Supply (in Ready) was <80V or >210V c) Average output current >100A. If these are not the problem, then replace the PPBM.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues