Pular para o conteúdo principal
GE Healthcare · MRSigna22257482225734

Will Not Scan With RF Errors Due to Defective SRFD Amp.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2225748 Analogic RF Amplifier 1 Peak Power Fault Error: 2225748 Host: lx-mr Process: NSP File: HPC PORT ID:spi rf mon Line: 3066 Analogic RF Amplifier 1 Peak Power Fault. Peak power from Analogic RF Amplifier 1 exceeds specification(69.5-69.8dbm). Possible causes: RF Interface (RFI) calibration, defective RFI output stage, defective Exciter or defective RF Amplifier 1

2225734 RF Interface (RFI) Mismatch Fault. The Dump Resistor in the RFI has overheated. 2225734 "RF Interface (RFI) Mismatch Fault. The Dump Resistor in the RFI has overheated. Possible causes: improper calibration, missing or improper cables between RFI and amplifier(s), defective RFI or defective Analogic RF Amplifier(s).

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2224562-2 1.5T RF AMP POWER MODULE W/ FRU PKG (SRFD)
Area
MR