GE Healthcare · CTLightSpeed182614182607260001022260001008260001012260001007
LightSpeed 4.X - GRE
Symptom
Scan aborts WITH Communications with the CCB sub-system has not been established. Scan aborts WITH Communications with the CCB sub-system has not been established
Error message
Error: 182614 Tilt drive is not stopping within 1500 msec. Expected stopping position XX (.01 deg) Current position YY (.01 deg). See Long Description for error details and example Host: ETC Process: Tilt File: tmcShutDown.c 1.31 1/29/93 Line: 121 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 5478/4/0 Type: None/Unknown Exception Level: Pri/Soft Time: 16:25:56:115 Log Series: 151 HW/SW error: Tilt drive is NOT stopping within 1500 msec. Expected Stopping Position -50 (.01 deg). Current Position -23 (.01 deg). Tilt Direction: 0 (0 = BCK 1 = FWD). Error: 182607 Host: ETC Process: Tilt File: CmdMotion.c 1.21 2/17/95 Line: 139 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 7839/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 12:12:01:160 Log Series: 617 HW/SW error: Velocity tracking error. Actual Velocity 0 (.01 deg/sec) Expected Velocity 100 (.01 deg/sec) Tol 80 (.01 deg/sec) Cur Pos -2181 (.01 deg) Tilt Dir: 1 (0 = BCK 1 = FWD). INFO: Probable Problem could be 180V Power is LOW. Error: 260001022 Communications between the DCB sub-system and OBC have been lost NOTE: Error: 260001022 can also state CCB to OBC comm problem but is a different err
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5191271 MDAS 16 Fuse Holder Upgrade Kit. Supercedes part number 46-170042P28. See Long Description for more information. Part #: 5191271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: MDAS 16 Fuse Holder Upgrade Kit FRU: Yes Status: Orderable Comments: Per ECR 2036788, 212008898, 11/30/2006. MDAS Fuse Holder Replacement Kit Contents: Part #: 46-170042P28 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: 30A, 250V PANEL-MOUNT FUSEHOLDER. FRU: Yes Status: Orderable Comments: FLS - FUSE HOLDER PART OF FMI KIT ** FE Assist ** IF FUSE HOLDER P/N 46-170042P28 FAILED, PLEASE ORDER MDAS FUSE FIX ON FAILURE KIT 5191271 (12/12/06 212001261)
- Area
- CT
Full cause & step-by-step resolution