GE Healthcare · MRSigna2228303
Artifact On Pure Study Due To Extending The Calibration Scan Beyond The Anatomy to be Scanned.
Symptom
MRIQ Some Images in Series Appear To Have Failed to Reconstruct Just a Blank image Some Images in Series Appear To Have Failed to Reconstruct Just a Blank image.
Error message
2228303 Part of scan region is outside the PURE calibration volume. Images will be blank where the scan does not overlap the calibration.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution