Pular para o conteúdo principal
GE Healthcare · MRSigna11002200002356

Unable to Communicate with the Fuji Synapse PACS due to Customer's Router Was Powered Down.

Symptom

Unable to Communicate with the Fuji Synapse PACS See Long Description Suddenly unable to communicate from the system to and from the Fuji Synapse PACS. Cannot transfer images either way and cannot pull a worklist. The errors shown below are logged. Sun Apr 12 09:28:49 2009 9007 4GE_NNET_NTWK_IC0 netserver presentation.c 792 read stamp failed (25) Connection reset by peer Sun Apr 12 09:29:20 2009 11002 4GE_NNET_NTWK_IC0 dcserver dcscommon.c 228 DCM low level error: Network error (508): Network error (114) <--- (many of these) On system reboot... Sun Apr 12 11:24:23 2009 9001 4GE_NNET_NTWK_IC0 msgView application.c 260 Unable to set connection - server NETserver host GE_NNET_NTWK_IC0. Sun Apr 12 11:25:30 2009 200002356 GE_NNET_NTWK_IC0 hostMonitorssProcStop.c 887 Exception Class: UnknownFunction: Host: System Startupdsserver died unexpectedly, no action being taken. Sun Apr 12 11:29:53 2009 11002 4 GE_NNET_NTWK_IC0 dcserver dcscommon.c 228 DCM low level error: Network error (508): Network error (114) Sun Apr 12 12:17:51 2009 2219945 GE_NNET_NTWK_IC0 scn PPS_IF.c 391 PPS :: Failure to execute event: IN PROGRESS event in PPSproxy process for Exam: 5671 Also when trying to pull a worklist using ConnectPro, these messages are found logged in the /usr/g/hisris/wlsErrors log: WLServer.c (311) Error code: 28 Fatal Unknown DCM ERROR

Error message

11002 DCM low level error. Network error (508). Network error (114) Thu Sep 30 07:56:30 2004 Error: 11002 Host: chmr Process: dcserver File: dcscommon.c Line: 228 DCM low level error: Network error (508): Network error (114)

200002356 dsserver died unexpectedly, no action being taken. Mon Nov 22 10:46:20 2004 Error: 200002356 Host: MRI1-OC0 Process: hostMonitor File: ssProcStop.c Line: 906 dsserver died unexpectedly, no action being taken.

9007 read stamp failed 25 Connection reset by peer See Long Description 9007 GE_NNET_NTWK_IC0 netserverpresentation.c 792 read stamp failed (25) Connection reset by peer

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution