GE Healthcare · CT
CT/e, CT Prospeed AI/FI Irix or Linux
Symptom
Scanning aborted with tube exceeded 70deg C .
Error message
Error: 110-0903H Tube exceeded 70°C See LD for error explanation Error code: 110-0903H Message/explanation: Tube exceeded 70°C Tube temperature switch detected open Potential cause: X-ray tube too hot; normal error Cooling problem Wiring problem Sensor problem (Tube) Rotation board Recommended action/ Troubleshooting guide: Wait until tube cools down for error clearance. Troubleshoot tube cooling system (if applicable) If error persists, short circuit the sensor feedback (between pins 15 and 16 on rotation board connector J2) If error disappears check cabling to tube and tube thermal switch state. If no, replace rotation board
Applies to
- Product
- CT/e, CT Prospeed AI/FI Irix or Linux See long description for sale information Product Name: CT/e Single CT Scanner: Irix Platform A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) 3.01 / 3.02 / 3.03 Linux CTe Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found in the attachment Always use the most appropriate documentation for your system.
- Component
- D3161T / D3162T Venus tube used on CTE (D3161T is NEW non reload version) See Long Description for details X-ray tube D3162T / D3161T , Modified Venus tube Part #: D3162T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: MX135 CT TH.1.1 RELOAD FRU: Yes Status: Orderable Comments: FLS - ECONOMY SLIP RING CT TUBE (ALSO CALLED CTE HIGHSPEED) 23 OHM (STATOR RESISTOR VALUE) 11/24/99 AM SOURCE = MEXICO FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA Part #: D3161T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: MX135 CT TH.1.1 GREEN FRU: Yes Status: Orderable Comments: This is GREEN tube, order only if New is required , If not please SEE D3162T, ms13452
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.