Functool Crashes While Creating SSAVE Report due to a Software Problem.
Symptom
Functool Crashes While Creating SSAVE Report See Long Description iTrak # 13265419 SPR # MRIhc44151 Functool crashes while creating SSAVE report. While post-processing multiple exams, Functool crashes while generating SSAVE report. If Functool is re launched, everything is fine. Generally, every time Functool loads a new exam, Functool allocates memory to maintain an array of image locations. While creating a SSAVE Report, a pointer is being used to point to the current image location. When Functool process is launched, the pointer to the current image location(used during SSAVE report) gets assigned correctly. Because of this, when SSAVE report is created, Functool doesn't crash. After selecting another data set in the browser, if Functool is opened with this new data set (without closing Functool), then the pointer used during SSAVE does not get assigned correctly. Due to this assignment, when SSAVE report is created, the Functool crashes. The root cause found to be a coding error.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.