GE Healthcare · X-ray R&FOptimaDiscovery10906171090453640001612207931220733
System experiences a shutdown of the detector power supply (DPS) during normal operation
Symptom
System experiences a shutdown of the detector power supply (DPS) during normal operation See Long Description and errors are displayed on the console. This problem is particularly prevalent when using or immediately after using the Table Rear TRAD Port. Performing tube angulation, OTS vertical travel or column rotation can also contribute to this failure mode due to an increase in CAN bus load.
Error message
ImChain Not Alive 1090617 ImChain Not Alive · 1090617 OTS CAN Command 1090453 OTS CAN Command · 1090453 Device Comm 6400016 Device Comm · 6400016 Detector Power Supply (DPS) 1220793 and 1220733 Detector Power Supply (DPS) · 1220793 · 1220733
Applies to
- Models
- Optima, Discovery
- Product
- Optima XR640 Discovery XR650
- Component
- 5192765 – ALB Cable Harness 5192765 – ALB Cable Harness 5251763 – ALB I/F Board 5251763 – ALB I/F Board
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- Definium 8000 Wallstand or Table Only Configurations running VolumeRAD M3 Software.
- Definium 8000 OTS dead - multiple long and can errors 1090451, 1090453, 1090440, 1090487, 1090489 due to Long Drive Asm failure.
- Collimator light does not work from the RCIM2 even though the button is present
- OTS drops vertically when using Auto Position
- Discovery XR650 or Optima XR640
- No procedure to copy IP looks from a Definium 8000 system to a Discovery XR650 or Optima 640 system