Export Function of Proto Exchange Never Finishes due to a Software Problem.

Symptom

Export Function of Proto Exchange Never Finishes See Long Description When exporting 25 site protocols using the Proto Exchange tool to CD/DVD it takes more than one hour with no sign that the CD/DVD is being accessed. It never finishes and finally the operator gives up on it. There are no error messages seen or logged. But if only a few protocols are being exported (I checked 3 protocols), the process takes only 3 to 5 minutes, and completes successfully. iTrak # 13260557, SPR # MRIhc44432.

Error message

No space left on device Error is displayed on screen during ProtoExchange session.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR