GE Healthcare · MRSigna
Export Function of Proto Exchange Never Finishes due to a Software Problem.
Symptom
Export Function of Proto Exchange Never Finishes See Long Description When exporting 25 site protocols using the Proto Exchange tool to CD/DVD it takes more than one hour with no sign that the CD/DVD is being accessed. It never finishes and finally the operator gives up on it. There are no error messages seen or logged. But if only a few protocols are being exported (I checked 3 protocols), the process takes only 3 to 5 minutes, and completes successfully. iTrak # 13260557, SPR # MRIhc44432.
Error message
No space left on device Error is displayed on screen during ProtoExchange session.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.