Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260134629859770550002750049398

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Table Encoder wire is broken. See long description. Table Encoder wire is broken. A foreign object got caught under the Table Cradle. The Encoder wire was broken trying to free the object.

Error message

Error: 260134629 GTCB Cradle absolute encoder malfunction during motion current p see long description for full error Error: 260134629 Host: tgp Process: tgp subsys GTCB Cradle absolute encoder malfunction Mon Mar 9 09:08:09 2009 Error: 260134629 Host: tgp Process: tgp subsys GTCB Cradle absolute encoder malfunction during motion current pos (linear) = 859770 target speed = 55000 um/s expected move = 27500 um actual move = 49398 um (abs)

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5127502 CRADLE WIRE ASSY Positioing GT-2000mm table only See long description. 5127502 CRADLE WIRE ASSY Positioing GT 2000mm table only Part #: 5127502 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CRADLE WIRE ASSY Positioing GT-2000mm table only FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues