Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230017082394680230023046

LightSpeed 5.X (60kW)

Symptom

Ring, Spot, or Band artifact on every 2nd, 4th, 8th image, or a similar pattern. See Long Description for more information. Ring or Spot artifact on every 2nd, 4th, 8th image or a similar pattern, depending on CU using 2i, 4i, 8i images, etc. Often the pattern will be one image out of every group of images, but may be a number of consecutive images in each group of images. images 2, 6, 10 in a 4i scan. or images 1,2,3,17,18,19 in a 16i scan.

Error message

Error: 230017082 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.3 The row number can be VERY confusing. Please see Long Description for how to decode the ROW NUMBER Tue Mar 22 20:12:04 2005 Error: 230017082 Host: ct01 Process: accs File: aircal_processing.c Line: 917 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.394680 Click Here to Decode the ROW NUMBER

Error: 230023046 Calibration limit check failed. Beam may be obstructed See LD for complete error Tue Mar 22 20:12:04 2005 Error: 230023046 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3953 Calibration limit check failed. Beam may be obstructed

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2355552 MDAS 16 Convertor Card ( replaces 2322994-2 ) Used In LightSpeed 16 Gantry with MDAS. 2355552 MDAS 16 Converter Card used in LS 4X with 16 slice MDAS (replaces 2322994-2) Part #: 2355552 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.25 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: SERVICE, MDAS16 CONV BD FRU: Yes Status: Orderable Comments: Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money. This is an MDAS16 CONV BD .compared to the SDAS. Make sure the correct Convertor is Ordered and replaced.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues