Symptom

Scanner hardware stopped scan with Communication To The Detector Heater Control Board Has Not Been Established See long Description for more details Scanner hardware stopped scan because of a Communication To The Detector Heater Control Board Has Not Been Established - Error 260132516 and 260132493 DCB to DHC (via CFC) communications failure. Communication To The Detector Heater Control Board Has Not Been Established - Error 260132516. Durring the hardware reset the DCB verifies that the communication to the Detector Heater control board can be properly established.

Error message

Error: 260132901 DCB to CFC communications failure See Long Description Mon Jun 29 07:52:19 2009 Error: 260132901 Host: orp Process: orp subsys File: ThermalManager.cxx Line: 1981 Mon Jun 29 07:52:18 2009 Host : Das Ermes # : 260132901 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 1981 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 DCB to CFC communications failure Failure code: -3

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5150070 or 5212666 DHC Detector Heater Controller - used on VCT and HD750 See LD for details 5150070 is non-RHOS complaint 5212666 is RHOS Complaint Part #: 5212666 RHOS compliant Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: DHC Assembly FRU: Yes Status: Orderable Part #: 5150070 non-RHOS compliant Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: DHC ASM FRU: Yes Status: Non Orderable <as of July 4, 2006 Comments: Added per ECR 2030869 6/26/2006, MS11892. DHC Detector Heater Controller
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.