No receive signal from any port except A port Due to 16 Channel Switch Cables Swapped.

Symptom

No receive signal from any port except A port see long description Only coils plugged into port A of 16 channel system had signal.NO errors except aps failed 2224666.

Error message

2247384 Diagnostic error. The signal's energy level is severely below the what is expected. See Long Description 2247384 tempmr diagCoord rrfRxFruIso.c 1539 Diagnostic: 16 Channel Switch Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 18571 Object: APS Test Task Test: '16 Ch Switch Input Osc. AB, Port A' diagnostic error. The signal's energy level is severely below the what is expected. 'Dead' channels are: 1-8-> Extended Error: Sys Cab and UTNS Bypass Diag: Dead channels while running these diags indicate a problem with a board in the receive chain (MUX, UTNS, Rx), or cables between the boards. Sixteen bad channels (or eight bad channels in an eight channel system would indicate a problem with the signal source, or problems other than the three receive chain boards. Eight bad channels in a 16 channel system would indicate a bad board. Swapping good boards with suspected boards is an effective way to isolate a problem. Problems with banks of channels (1-4, 5-8, etc) may indicate a problem with a receive When individual bad channels (1,8,9,16) may lead to an improperly seated cable.\n 16 Channel Switch diag: If problems exist when running this diagnostic, the Sys Cab Loopback d

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.