Pular para o conteúdo principal
GE Healthcare · MRSigna2224842

When the Cradle Returns to Home, it Will Reverse Direction and Move Toward the Bore of the Magnet.due to Shorted Pen Pan…

When the Cradle Returns to Home, it Will Reverse Direction and Move Toward the Bore of the Magnet.due to Shorted Pen Panel Filter (at PP1-J47) for Cable Run 387.

Symptom

When the Cradle Returns to Home, it Will Reverse Direction and Move Toward the Bore of the Magnet.

Error message

2224842 SRI Hardware Error Unexpected cradle motion. Thu Jun 9 13:59:11 2005 Error: 2224842 Host: MR_PHC Process: NSP File: HPC_PORT_ID:sri Line: 99 SRI Hardware Error: Unexpected cradle motion. This error would occur if the table were moved by hand. Please run SRI diagnostics if that is not the case. Expected Cradle Velocity: -100 mm/sec Measured Cradle Velocity: 189 mm/sec

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
46-271920P3 Filter Assembly See Long Description 46-271920P3 Filter Assembly 25-PIN SUB-D MALE-FEMALE FILTER ASM (This is the pen panel filter that the table elevation (dock) motor drive cable Run 387 connects to, to J47 on the pen panel). In FEMC this part is listed as "CONNECTOR".
Area
MR

Full cause & step-by-step resolution