VG gantry Error #75 Drivers Voltage Test H.W.D.T Activated Voltage Disconnected.

Symptom

VG gantry Error #75 Drivers Voltage Test H.W.D.T Activated Voltage Disconnected. VG hawkeye Error #75 happens intermittently. Sometimes switch off the power to the gantry for 10 minutes and then switch it on and all is Ok.

Error message

Error #75 (Error 75) Drivers Voltage Test H.W.D.T Activated Voltage Disconnected GANTRY_ERROR_HWDT_VLTG

Applies to

Models
Millennium
Product
VGH VG and VG Hawkey and Varicam
Component
5197979 (PCA000492) MCH Board for SPARE PART 5197979 (PCA000492) MCH Board for SPARE PART 5197979 MCH Board is not compatable with VG motion software version 6.0AH Verify VG Gantry Firmware version. VG Gantry Firmware Version must be 6.0P7 . See KOID 153674
Area
Nuclear Medicine