Internal failure of the Lytron chiller caused it to chill the coolant to 25F.

Symptom

Bore is cold to the touch and has condensation on it. See Long Description /usr/g/service/log/pcft.log shows bore temps of 8 C overnight. FE found ice on the gradient coolant lines.

Error message

2225542 Dynamic disable fault, Rear of Coil short. Dynamic Disable bias fault on back drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on BACK of BRM. 2225542 "Dynamic disable fault, Rear of Coil shortDynamic Disable bias fault on back drivers. Possibly a short circuit loador a shorted diode in the dynamic disable switch boxes on BACK of BRM."

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2222564-30 50 and 60 HZ chiller by Lytron. See Long Description 2222564-30 50 and 60 HZ chiller by Lytron Model Number RC045J03BG0C011 Specified by 2222564PSP. Lower level FRU's should be utilized before attempting to replace the entire chiller. Gradient Chiller Motor - 2222564-25 Pump Assembly for Lytron Chiller - 2222564-29
Area
MR

Full cause & step-by-step resolution