Application Software does not complete the boot process with TPS Download failure due to poor connection of coaxial cabl…

Application Software does not complete the boot process with TPS Download failure due to poor connection of coaxial cable at J5 LINE TRIGGER at rear of TPS/ISE Fan/Power Module.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2226595 Unable to measure a valid line frequency See Long Description 2226595 MRSC_OC0 NSP IPG:line_freq_check.c 184 IPG INIT: Unable to measure a valid line frequency; thus, cannot verify 'lineFreq' value in /w/config/MRconfig.cfg. Make sure this value is set correctly. Run LCA-related IPG diags.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.