Pular para o conteúdo principal
GE Healthcare · MRSigna2218523

System Hang or Crash Downloading Protocol Due to Corrupt Protocol.

Symptom

Software Crashes to Wallpaper and 6 Icons Software Crashes to Wallpaper and 6 Icons. Icons are located at top left corner of screen.

Error message

2218523 Internal software error Current prescription has no PSD Thu Jul 6 10:08:46 2006 Error: 2218523 Host: MRI_HDe Process: scn File: MsaScanMgr.m Line: 1203 Internal software error Current prescription has no PSD

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR