LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry

Symptom

Special Fast Cals pops up everyday even after running them; intermittent ring artifacts & errors in calculations during Special Cals Special Fast Cals pops up everyday even after running them; intermittent ring artifacts & air cal errors in calculations

Error message

Error: 230023078 Air Ratio Calibration was NOT completed successfully See long description for details. Sat Jul 18 09:41:32 2009 Error: 230023078 Host: ct99 Process: dailyPrepRx File: StateMachineActions.c Line: 2119 Air Ratio Calibration was NOT completed successfully

Error: 230017186 ARCal K or B exceeds limit See long description for details Sat Jul 18 09:42:08 2009 Error: 230017186 Host: ct99 Process: accs File: AirRatioCal.c Line: 2119 ARCal K or B exceeds limit Scan number=2, Row=14, Channel=368 K=-0.818094 B=1.027403

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.