Pular para o conteúdo principal
GE Healthcare · CTLightSpeedHiSpeed18290050412182902

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Table will not move. Table will not move

Error message

Error: 182900 Stopping scan because a tableside emergency off button has been pressed Open Long Description to see example error Sun Jul 18 10:02:54 2004 Error: 182900 Host: ETC Process: Table Controls File: tc_scan.c 1.14 1/11/94 Line: 235 Function: No System Function Reported Scan: 50412/1/3 Type: Axial Exception Level: Pri/Most Severe Time: 10:02:54:710 Log Series: 57 Stopping scan because a tableside emergency off button has been pressed

Error: 182902 Emergency Off button has been pressed Sun Sep 5 16:04:39 2004 Error: 182902 Suite: RP21 Host: ETC Process: Table Controls File: tc emergency.c 1.6 2/16/99 Method: No Method line: 89 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 4016 Log Series: 22 Emergency Off button has been pressed

Applies to

Models
LightSpeed, HiSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-297698P3 27" HIGH PROFILE Tape switch See the LD for tape switch jpg Part #: 46-297698P3 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 27" HIGH PROFILE BEAD, RIBBON SWITCH FRU: Yes Status: Orderable TABLE TOP COVER TAPE SWITCH. FRU = 2. *P3 WILL REPLACE THE P2 WHEN P2 STOCK IS DEPLETED. FE: IF YOU ARE REPLACING 46-297698P2 YOU MUST ORDER BOTH 2216522 AND 46-297698P3 46-297805G2 Table Cover Black ByPass JUMPER CONNECTOR See LD for details and to make sure correct part ordered Part #: 46-297805G2 Part Type: GP CONSUMABLES Part Weight: 0.04 LBR Repairable: No Returnable: No Hazardous Material: N Description: JUMPER CONN. FRU: Yes Status: Orderable Comments: FLS - JUMPER CONNECTOR ADDED PER PCN 220369. USED ON CT TABLE ASSEMBLY. CHANGING FROM FRU=N TO FRU=1 PER SAME PCN. MSA8429 01/30/02 See differences between G1 and G2 jumpers below. Wiring also shown for the two style jumpers. 2216522 CHANNEL EXTRUSION. See the LD for part comments Part #: 2216522 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: CHANNEL EXTRUSION FOR TAPE SWITCH FRU: Yes Status: Orderable CHANNEL EXTRUSION FOR TAPE SWITCH(46-297698P3). FRU = 2 THIS PART MUST ALWAYS BE ORDERED WHEN PART# 46-297698P3 IS ORDERED . 46-297805G1 Table ByPass JUMPER CONN See LD for details and to make sure correct part number ordered 46-297805G1 Table BYPASS JUMPER CONN Part #: 46-297805G1 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: JUMPER CONN. FRU: Yes Status: Orderable See differences between G1 and G2 jumpers below. Wiring also shown for the two style jumpers.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.