GE Healthcare · CTLightSpeed200287513200280006
LightSpeed 4.X - Octane
Symptom
System will not AutoFilm. Filming done manually. System will not AutoFilm. Filming done manually System will not Auto Film. Filming done manually
Error message
Error: 200287513 The request failed to display for exam: ### See LD for complete error Wed Jan 31 17:36:43 2007 Error: 200287513 Host: lhcct2 Process: afViewer File: /gspnet/blr/component/commondisplay/view/ Line: 2676 AutoFilmCntl::displayFirstImage() The request failed to display for exam: 4 952 series: 2 images: 32 type: AXIAL Error: 200280006 Null pointer detected See Long description for more details Sat Jan 4 10:58:53 2003 Error: 200280006 Host: ct02 Process: xrViewer File: /tmp_mnt/homes/adwaitm/H16_mke/ commondisp Line: 276 Null pointer detected
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 405I.2_H4.0M5 for LightSpeed 4.X (LightSpeed 16) with H16 Console (GOC1)(2341923, 2341923-2) - showprods returns 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395173 1204I.2_ H4.0M5 CLASSM LightSpeed/ HiSpeed QX/I Class M Software 2395176 2395170, 2395171, 2399867 LightSpeed 16 Applications Software (Disk 1, Disk 2 and Patch A) 405I.2_H4.0M5 FW47-03 LS2002M5. IRIX only. Class C/M Common w/ LS 2X, 3X, & 4X.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required