GE Healthcare · MRSigna2225622
The system failed to scan PROBE withe error 2225622.
Symptom
Unable to Scan PROBE. This an LX system running on 9.1 software with an Erbtec amplifier, the system failed to scan PROBE with error: 2225622
Error message
2225622 ERBTEC RF Amplifier Fault 32, Unblank longer Than 20 msec. Mon Mar 9 20:13:09 1998 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225622 File: PMH.m 1.29 Method: spi_rf_mon Line: 3035 ERBTEC RF Amplifier fault 32. Unblank longer than 20 msec. The rising and falling edge of the unblank signal did not fall within the 20 msec time limit. Check the unblank cable and the Processor Board.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution