Grafidy will not run on an excite-based system if it is configured as a high speed with the hispeed key installed.

Symptom

Grafidy long terms will not run. A pop up error comes up stating java.lang.exception:read measured data (linear and b0) error for long xx Raw file indicates this Vlong : XX See Long Description This popup message is not logged on the system.

Error message

java.lang.exception:read measured data (linear and b0) error for long xx Raw file indicates this Vlong : XX See Long Description This is a pop up error message that is not logged to any error log.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR