Symptom

The Main Circuit Breaker in the Teal PDU in the HFD Cabinet will not reset The main circuit breaker in the Teal PDU in the HFD Cabinet will not reset. It will not even mechanically reset (i.e. stay in the engaged position) with all power off. It doesn't "catch", it immediately swings back to the tripped position, even with all power off. With power off it should "lock" into the on position, and stay there, at least until power is applied. Therefore this is not an electrical problem.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5118444 Includes cabinet and PDU, must swap rest of cabinet over. 5118444 Includes cabinet and PDU, FE must swap rest of cabinet over.